Part No: MICROWAVEMW800 |
| |
The Transform 800
top-loading closed vessel Microwave offers a high-pressure system to enhance
digestion quality while decreasing digestion times. This system is capable of
simultaneously running up to 10 high-pressure closed vessels, providing a fast,
automated method to digest even the most difficult samples.
With simple
one-step operation and full computer control of temperature, pressure and
power, this cost-effective MW 800 consolidates your AA, ICP-AES, and ICP-MS
applications. |
| |
| |
 |
 |
Round, pressure resistant heavy duty oven chamber made of PFA-coated stainless
steel |
|
 |
Evenly distributed microwave field throughout the chamber to ensure uniform
heat for each sample vessel |
|
 |
Closed vessels designed to operate at up to 250°C temperature and 800psi
pressure with up to 10 vessels |
|
 |
Direct sensing and controlling of sample temperature and pressure of closed
vessels to ensure reproducible digestion every time |
|
 |
On-line EPA protocols and methods directory, full compliance with EPA methods
|
|
 |
Real-time, Microsoft windows compatible software for remote control via
external PC |
|
| |
| |
|
| The Transform 800 can
hold 10 high-pressure vessels capable of digesting at 800 psi and 250°C. A
sealed thermocouple and pressure sensor enable fast and accurate temperature
and pressure control. The MW 800 microwave system is available at less cost
than comparable microwave systems but with equivalent safety and performance
ability |
| |
| |
|
|
 |
| |
 |
 |
| |
 |
| The Transform 800
comes standard with a range of safety features: |
| |
 |
Fume exhaust module |
|
 |
Over pressure vent detection mechanism shuts down the microwave power
applicable for each vessel |
|
 |
Digestion vessels molded with engineering plastics |
|
 |
Over-pressure ruptured disks on each vessel |
|
 |
Continuous temperature and pressure monitoring and control |
|
 |
Release of acid fumes into collection reservoir in vessel carousel |
|
| |
| |
 |
| Transform 800 is a
full computer controlled system. Its software features real-time control of
temperature and pressure with graphic display, automatic data storage,
unlimited method storage, and complete record of instrument parameters. |
| |
| |
 |
| For full data sheets
and a written quotation please contact our Mindex office on
sales@mindex-ltd.co.uk including
your company name, telephone numbers and details of your product of
interest |
|