Part No: ATACD86 |
| |
The model D86 BOILAR
continuously measures the percent recovered boiling point temperatures of
petroleum products from 5%-95% in the temperature range 50-400 deg
C.
Simple to operate and maintain, the D86 boilar si the perfect choice
for single point boiling point analys |
| |
| |
A sample is metered
into the vaporizer is assembly at a constant flow.
The sample then
flows down a heater assembly maintained at a temperature dependant on the
percentage of sample to be boiled off the residue is collected in a measuring
device the sample is then discharged at rate proportional to the sample input
rate.
The ratio of the input pump to outlet pump will define the % of
sample recovered.
The temperature of the vaporizer at equilibrium is
the percent recovered boiling point. This temperature is measured and
transmitted. The sample metering arrangements are infinitely variable between
5% and 95%.
The percentage recovered boiling point can range from
50-400 deg C. |
| |
|
|
 |
| |
 |
 |
| |
| |
 |
 |
High speed continuous boiling point analysis |
|
 |
Self Diagnostic |
|
 |
Correlates to ASTM D86 and ISO 3405-IP123 and surpasses test method
repeatability |
|
 |
Accurate and reliable |
|
 |
Ideal for closed loop control of process plant |
|
 |
Temperature 50-400 deg C for 5% - 95% distilled |
|
| |
| |
| For full data sheets
and a written quotation please contact our Mindex office on
sales@mindex-ltd.co.uk including
your company name, telephone numbers and details of your product of
interest |
| |
|
|